Intravoxel incoherent motion (IVIM) imaging at different magnetic field strengths: what is feasible?

Rydhög AS, van Osch MJP, Lindgren E, Nilsson M, Lätt J, ahlberg FS, Wirestam R, Knutsson L. Intravoxel incoherent motion (IVIM) imaging at different magnetic field strengths: what is feasible?. Magn Reson Imaging. 2014;32(10):1247–58.

Abstract

BACKGROUND: Due to limited SNR the cerebral applications of the intravoxel incoherent motion (IVIM) concept have been sparse. MRI hardware developments have resulted in improved SNR and this may justify a reassessment of IVIM imaging for non-invasive quantification of the cerebral blood volume (CBV) as a first step toward determining the optimal field strength. PURPOSE: To investigate intravoxel incoherent motion imaging for its potential to assess cerebral blood volume (CBV) at three different MRI field strengths. MATERIALS AND METHODS: Four volunteers were scanned twice at 1.5 T, 3 T as well as 7 T. By correcting for field-strength-dependent effects of relaxation, estimates of corrected CBV (cCBV) were obtained in deep gray matter (DGM), frontal gray matter (FGM) and frontal white matter (FWM), using Bayesian analysis. In addition, simulations were performed to facilitate the interpretation of experimental data. RESULTS: In DGM, FGM and FWM we obtained cCBV estimates of 2.2 ml/100 ml, 2.7 ml/100 ml, 1.4 ml/100 ml at 1.5 T; 3.7 ml/100 ml, 5.0 ml/100 ml, 3.2 ml/100 ml at 3 T and 15.5 ml/100 ml, 20.3 ml/100 ml, 7.0 ml/100 ml at 7 T.
Last updated on 02/26/2023